CapnOxi meditech Capnograph Calibrating CO2 Measurement It is recommended to calibrate CO2 measurement when you start CO2 monitoring
CapnOxi meditech Capnograph

Calibrating CO2 Measurement

Calibrating CO2 Measurement
It is recommended to calibrate CO2 measurement when you start CO2 monitoring. But you must carry out CO2 calibration when CO2 needs zero calibration displays in alarm area.
Procedure:
1. Press menu key on keyboard to enter Main Menu dialog.
2. Select User Settings button to enter User Settings dialog.
3. Select Calibrate Zero button to calibrate CO2 measurement.
4. If 25 seconds later, if calibration failed, failure message will be shown.